Elliot Lake Research Field Station of Laurentian University

Analytical Services - QA\QC

Home Laboratory Contact List Publications History

 

 

 

 

 

 

 

 

 

 

QUALITY ASSURANCE / QUALITY CONTROL

To ensure the best results for our clients, the Analytical Services Laboratory has implemented a comprehensive Quality Control/Quality Assurance Program.   Quality Assurance is a set of operating principles that, if strictly followed during the sample collection and analysis, will produce accurate data of defensible quality. This program is designed to comply with Canadian federal and provincial regulatory agencies and applies to all samples submitted to the laboratory. A comprehensive Quality Control/Quality Assurance Program is in place in the laboratory.

Elements of the QA/QC program include:

Staff organization, responsibilities and training requirements

Sample control and documentation procedures

Standard operating procedures (SOP) and method audits

Internal quality control activities and corrective action procedures

Equipment calibration and preventive maintenance procedures

Data quality assessment and reporting system

Currently, the QA/QC program is prepared and overseen by the Laboratory Manager, QC Chemist and the Scientific Advisor.

Overall QC protocols used include:

Control Standards: The stability of the instrument calibration is checked during the run by the analysis of the control standards.

Control Samples: The accuracy of the instrument calibration, recovery during sample preparation and long-term stability are checked by the analysis of control samples with each batch.

Blanks: Method blanks are analyzed to confirm the absence of contamination.

Duplicates: Duplicates are analyzed to check method and stability and performance.

Matrix Spikes: Sample matrix effects are checked by spiking samples with a known amount of standard and subsequently measuring the recovery of the analyte in the sample.

Standard Reference Materials: Certified standards are used as control samples to check the method performance in the matrix, but only when appropriate.

Laboratory bias is determined by a spike recovery or by duplicate analysis. As part of the QA/QC program, the QC Chemist and Laboratory Manager regularly submit duplicates and spiked reference samples to the laboratory technicians for analysis and then review the results to identify and rectify any identified laboratory biases.

Data assessment is conducted every 6 months for regular analytical programs, and as an integral part of any special projects. Quality assurance audits, such as inter-comparison studies, spike recoveries, quality control samples, and replication samples are used to maintain quality analytical results. Our QA/QC program has been established for our customers, to ensure them that their results are accurate and precise.

Performance Assessment

The Analytical Services Laboratory participates in CAEAL performance assessment programs (twice a year for soil and twice a year for water) to continuously monitor the performance of the methods used.  

Methods

Most routine analytical methods are based on established standard methods (e.g. APHA, AECB, Environment Canada). All routine methods are thoroughly validated and authorized before use in the laboratory. The methods are reviewed annually.

Instrument Calibration and Maintenance

All instruments are calibrated before sample analysis. The instrument response must be within specified limits. Also, to ensure rapid turn around, the instruments are properly maintained.

Elements of the Analytical Laboratory QA/QC Program

Element Frequency Limits
Containers Pre-cleaned not detectable
Recovery of known addition 10% of samples or 1 per batch plus or minus 10%
Reagent blank 10% of samples or 1 per batch <MDL
Traveling blank Project specific <MDL
Duplicate analysis 10% of samples or 1 per batch plus or minus 10%
Internal standards All samples within control limits
Certified Reference Standards All solid samples and metals CCRMP limits
Control charts Key parameters (such as Ra_226, Po_210, Th, U & TKN) plus or minus 2 SD (Standard deviation)
Data trends All routine samples ENVISTA warning limits
Internal QC check Monthly in-house limits
Method audits Yearly adhere to SOP
Staff performance audits Monthly  
Instrument Calibrations:  

Once a month

Each day

 
Balance
pH meter Twice a day 100g internal standard & check with 100mg std weight
UV Spectrometer Multi-point for each batch, verify every ten samples buffer, 2, 4, 7 & 10
Ion Chromatographer Multi-point daily >.99
AA Multi-point daily, verify every ten samples >.99 plus or minus 10% initial
Alpha, Beta & Gamma Spectrometer Multi-point

Once a year

>.99
ISE/Conductivity meter Every four hours  
Dissolved oxygen meter Daily slope within 54 to 60 mV
Micro-pipets (<5 ml) Monthly within manufacture limits

Comparison of Results

When it is known that the analyte levels for a particular sample site remain consistent over time, the laboratory compares each result to previous ones. Any changes are checked and if the change is confirmed, the client is notified.

Reporting Quality Control Data

Quality Control data can be reported at no additional charge.  A variety of formats are available to suit the client's needs.   Contact Al Douglas  (705) 675-1151 ext 1506 if this service is required.

 

    

Copyright © 2000
Last modified: June 26, 2006